Accuracy
without Option 4294A-1D5 ±20 ppm (at 23 ±5 °C)
±40 ppm (at 0 to 55 °C)
with Option 4294A-1D5 ±0.13 ppm (at 0 to 55 °C)
Voltage Signal Level
Range 5 mVrms to 1 Vrms
Resolution 1 mV
Accuracy
at four-terminal pair port of ±[(10 + 0.05 f)% + 1 mV] (at 23 ±5 °C)
the 4294A or 7-mm port of
the 42942A
at measurement port of the ±[(15 + 0.1 f)% + 1 mV] (at 23 ±5 °C)
42941A, 16048G/H
NOTE
f: frequency [MHz].
These characteristics apply when OPEN is connected to each port.
Test signal level should be 0.5 Vrms when the measured impedance is 50 .
Beyond 23 ±5 °C of temperature, test signal level setting accuracy is twice as
bad as described.
Current Signal Level
Range 200 µArms to 20 mArms
Resolution 20 µA
Accuracy
at four-terminal pair port of
the 4294A
at 15 MHz +[10% + 50 µA], –[(10 + 0.2 f
2
)% + 50 µA]
(at 23 ±5 °C, typical)
at >15 MHz ±[(10 + 0.3 f)% + 50 µA]
(at 23 ±5 °C, typical)
at 7-mm port of the 42942A
at 5 MHz +[10% + 50 µA], –[(10 + 1 f
2
)% + 50 µA]
(at 23 ± 5 °C, typical)
at >5 MHz ±[(10 + 0.3 f)% + 50 µA]
(at 23 ±5 °C, typical)