3
at measurement port of the
42941A, 16048G/H
at 5 MHz +[10% + 50 µA], –[(15 + 1.5 f
2
)% + 50 µA]
(at 23 ±5 °C, typical)
at >5 MHz ±[(20 + 0.3 f)% + 50 µA]
(at 23 ±5 °C, typical)
NOTE
f: frequency [MHz].
These characteristics apply when SHORT is connected to each port.
Test signal level should be 20 mArms when the measured impedance is 50 .
Signal Level Monitor
Voltage range (Same as the voltage signal level setting range)
Voltage monitor accuracy
at four-terminal pair port of ±(10 + 0.05 f + 100/Z
X
)[%]
the 4294A or 7-mm port of (at 23 ±5 °C, typical)
the 42942A
at measurement port of the ±(10 + 0.15 f + 100/Z
X
)[%]
42941A, 16048G/H (at 23 ±5 °C, typical)
Current range (Same as the current signal level setting range)
Current monitor accuracy
at four-terminal pair port of ±(10 + 0.3 f + Z
X
/100)[%]
the 4294A or 7-mm port of (at 23 ±5 °C, typical)
the 42942A
at measurement port of the ±(10 + 0.4 f + Z
X
/100)[%]
42941A, 16048G/H (at 23 ±5 °C, typical)
NOTE
f: frequency [MHz], Z
X
impedance measurement value [].
Beyond 23 ±5 °C, the test signal level monitor accuracy is twice as bad as
described.
Output Impedance
Output impedance 25 (nominal)