Measurement Accuracy
Conditions of Accuracy Specifications
Temperature
Four-terminal pair port of the 23 ±5 °C
4294As front panel Beyond 23 ±5 °C, the measurement accuracy is twice as bad
as described.
7-mm port of the 42942A Within ±5 °C from the adapter setup temperature.
Terminal Adapter Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
3.5-mm port of the 42941A Within ±5 °C from the adapter setup temperature.
Impedance Probe Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Measurement terminal of the Within ±5 °C from the adapter setup temperature.
16048G or 16048H Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Measurement bandwidth 5
Measurement Accuracy
|Z|, |Y| accuracy ±E [%] (see Equation 1 on page 10, Equation 2 on page 12,
Equation 3 on page 14)
θ accuracy ±E /100 [rad]
L, C, X, B accuracy
at D
x
0.1 ±E [%]
at D
x
>0.1 ±E 1 + D
x
2
[%]
R accuracy
`b