4
DC Bias Function
DC voltage bias
Range 0 to ±40 V (see Figure 1)
Resolution 1 mV
Accuracy ±[0.1% + (5 + 30 |I
mon
|) mV](at 23 ±5 °C)
±[0.2% + (10 + 30 |I
mon
|) mV](beyond 23 ±5 °C)
DC current bias
Range 0 to ±100 mA (see Figure 1)
Resolution 40 A
Accuracy ±[2% + (0.2 + |V
mon
|/20) mA] (at 23 ±5 °C)
±[4% + (0.4 + |V
mon
|/20) mA] (beyond 23 ±5 °C)
DC voltage bias at constant
voltage mode
Range 0 to ±40 V (see Figure 1)
Resolution 1 mV
Accuracy ±[0.5% + (5 + Z
d
|I
mon
|) mV] (at 23 ±5 °C, typical)
±[1.0% + (10 + Z
d
|I
mon
|) mV] (beyond 23 ±5 °C, typical)
DC current bias at constant current mode
Range 0 to ±100 mA (see Figure 1)
Resolution 40 A
Accuracy ±[1% + (0.5 + |V
mon
|/10000) mA] (at 23 ±5 °C, typical)
±[2% + (1.0 + |V
mon
|/5000) mA] (beyond 23 ±5 °C, typical)
DC bias monitor
DC voltage range (Same as the dc voltage bias setting range)
DC voltage accuracy ±[0.2% + (5 + Z
d
|I
mon
|) mV] (at 23 ±5 °C)
±[0.4% + (10 + Z
d
|I
mon
|) mV] (beyond 23 ±5 °C)
DC current range (Same as the dc current bias setting range)