76 Programmer’s Guide
5 Command Reference
Amplitude Subsection
Reference level
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel: <value>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
This command sets the reference level for the Y-axis.
*RST: 0 dB
Range: –140.00 to +20.00 dBm
Example: :DISPlay:WINDow:TRACe:Y:RLEVel 10dBm
Key access: Amptd > Ref Level
Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator of the spectrum analyzer.
Range: 0 to 50 dB
Example: :POWer:ATTenuation 30dB
Key access: Amptd > Attenuation
Input Port Attenuator Auto State
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
This command turns on/off auto input port attenuator state.
*RST: On
Example: :POWer:ATTenuation:AUTO?
Key access: Amptd > Attenuation > Auto