114 Programmer’s Guide
5 Command Reference
Query Channel Power and Density Result
:MEASure:CHPower?
This command returns scalar results of main channel power, and power
density.
:MEASure:CHPower:CHPower?
This command returns the value of the channel power in amplitude
units.
:MEASure:CHPower:DENSity?
This command returns the value of the channel power density in
amplitude units/Hz.
Channel Power RRC Filter State
[:SENSe]:CHPower:RRC[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:RRC[:STATe]?
Use this command to turn on/off RRC filter in Channel Power test.
Example: :CHPower:RRC 1
Key access: Meas > More > RRC Filter
Channel Power RRC Filter Alpha Value
[:SENSe]:CHPower:RRC:ALPHa <val>
[:SENSe]:CHPower:RRC:ALPHa?
Use this command to set the Alpha value of RRC filter in channel power
test.
Range: 0.01 to 1
Example: :CHPower:RRC:ALPHa 0.5
Key access: Meas > More > RRC Filter