Agilent Technologies N9344C Manual de Serviço Página 146

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 224
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 145
138 Programmer’s Guide
5 Command Reference
CAT Center Frequency
[:SENSe]:CAT:FREQuency:CENTer <freq>
[:SENSe]:CAT:FREQuency:CENTer?
This command defines center frequency of cable and antenna
measurement.
Range: 5 MHz to 7 GHz
Example: :CAT:FREQuency:CENTer 200 MHz
Key access: Freq > Center Freq
CAT Frequency Span
[:SENSe]:CAT:FREQuency:SPAN <freq>
[:SENSe]:CAT:FREQuency:SPAN?
This command defines the frequency span of current measurement.
Example: ;CAT:FREQuency:SPAN 200 MHz
Key access: Span > Span
CAT Distance Start
[:SENSe]:CAT:DISTance:STARt <distance>
[:SENSe]:CAT:DISTance:STARt?
This command defines the start distance point of DTF measurement
Range: 0 to Stop distance - 0.1
Example: :CAT:DISTance:STARt 1.3
Key access: Meas > Start Distance
CAT Distance Stop
[:SENSe]:CAT:DISTance:STOP <distance>
[:SENSe]:CAT:DISTance:STOP?
This command defines the stop distance of DTF measurement
Range: 0 to stop distance – 0.1 meter
Example: :CAT:DISTance:STOP 100.3
Key access: Meas > Stop Distance