158 Programmer’s Guide
5 Command Reference
Set Frequency Step for Customer Scan Channel For Top N Sweep
:CHSCanner:EDIT:TOP:CUSTom:STEP <value>
:CHSCanner:EDIT:TOP:CUSTom:STEP?
This command specify the frequency step size of custom scan channels
for Top N sweep mode.
Example: :CHSCanner:EDIT:TOP:CUSTom:STEP 2000000
Key access: Meas > Range Edit > Edit Method Custom, then > Freq Step Size
Set Bandwidth For Custom Scan Channel For Top N Sweep
:CHSCanner:EDIT:TOP:CUSTom:BANDwidth <value>
:CHSCanner:EDIT:TOP:CUSTom:BANDwidth?
This command specify the bandwidth of custom scan channels for Top
N sweep mode.
Example: :CHSCanner:EDIT:TOP:CUSTom:BANDwidth 30000
Key access: Meas > Range Edit > Edit Method Custom, then > Bandwidth
Set Custom Scan Channel Quantity For Top N Sweep
:CHSCanner:EDIT:TOP:CUSTom:CHNumber <value>
:CHSCanner:EDIT:TOP:CUSTom:CHNumber?
This command defines custom scan channel quantity for Top N sweep
mode.
Example: :CHSCanner:EDIT:TOP:CUSTom:CHNumber 200
Key access: Meas > Range Edit > Edit Method Custom, then > Number of Chn
Set Edit Method For Bottom N Sweep
:CHSCanner:EDIT:BOTTom:METHod STD|CUSTom
:CHSCanner:EDIT:BOTTom:METHod?
This command specify the edit method for Bottom N sweep mode.
Example: :CHSCanner:EDIT:BOTTom:METHod CUST
Key access: Meas > Scan Mode > Bottom N > Return > Range Edit > Edit Method