Agilent Technologies N9344C Manual de Serviço Página 83

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 224
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 82
Command Reference 5
Programmer’s Guide 75
Full Frequency Span
[:SENSe]:FREQuency:SPAN:FULL
Set the frequency span to full scale.
*RST: 3.0 GHz
Example: :FREQuency:SPAN:FULL
Key access: Span > Full Span
Zero Span
[:SENSe]:FREQuency:SPAN:ZERO
Set the frequency span to zero span.
Key access: Span > Zero Span
Last Frequency Span
[:SENSe]:FREQuency:SPAN:PREVious
Set the frequency span to the previous span setting.
Key access: Span > Last Span
Auto Tune
[:SENSe]:FREQuency:TUNE:IMMediate
Auto tune the spectrum analyzer parameter to display the main signal.
Example: :FREQuency:TUNE:IMMediate
Key access: Freq > Auto Tune
Base Band Channel Switch
[:SENSe]:FREQuency:LFCHannel OFF|ON|0|1
[:SENSe]:FREQuency:LFCHannel?
Use this command to turn on/off low frequency channel.
Example: :FREQuency:LFCHannel ON
Key access: Freq > More > LowFreqChannel > ON/Off