Agilent Technologies N9344C Manual de Serviço Página 168

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 224
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 167
160 Programmer’s Guide
5 Command Reference
Set Start Frequency For Custom Scan Channel For Bottom N
Sweep
:CHSCanner:EDIT:BOTTom:CUSTom:STARt <value>
:CHSCanner:EDIT:BOTTom:CUSTom:STARt?
This command specify the start frequency of custom scan channels for
Bottom N sweep mode.
Example: :CHSCanner:EDIT:BOTTom:CUSTom:STARt 20000000
Key access: Meas > Range Edit > Edit Method Custom, then > Start Freq
Set Frequency Step for Customer Scan Channel For Bottom N
Sweep
:CHSCanner:EDIT:BOTTom:CUSTom:STEP <value>
:CHSCanner:EDIT:BOTTom:CUSTom:STEP?
This command specify the frequency step size of custom scan channels
for Bottom N sweep mode.
Example: :CHSCanner:EDIT:BOTTom:CUSTom:STEP 2000000
Key access: Meas > Range Edit > Edit Method Custom, then > Freq Step Size
Set Bandwidth For Custom Scan Channel For Bottom N Sweep
:CHSCanner:EDIT:BOTTom:CUSTom:BANDwidth <value>
:CHSCanner:EDIT:BOTTom:CUSTom:BANDwidth?
This command specify the bandwidth of custom scan channels for
Bottom N sweep mode.
Example: :CHSCanner:EDIT:BOTTom:CUSTom:BANDwidth 30000
Key access: Meas > Range Edit > Edit Method Custom, then > Bandwidth
Vista de página 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 223 224

Comentários a estes Manuais

Sem comentários