Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manual de Serviço Página 647

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 970
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 646
:TRIGger Commands 28
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 647
:TRIGger:LEVel:LOW
(see page 860)
Command Syntax
:TRIGger:LEVel:LOW <level>, <source>
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:LEVel:LOW command sets the low trigger voltage level voltage
for the specified source.
High and low trigger levels are used with runt triggers and rise/fall time
(transition) triggers.
Query Syntax
:TRIGger:LEVel:LOW? <source>
The :TRIGger:LEVel:LOW? query returns the low trigger voltage level for
the specified source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:LEVel:HIGH" on page 646
":TRIGger:RUNT Commands" on page 680
":TRIGger:TRANsition Commands" on page 691
":TRIGger[:EDGE]:SOURce" on page 661
Vista de página 646
1 2 ... 642 643 644 645 646 647 648 649 650 651 652 ... 969 970

Comentários a estes Manuais

Sem comentários