:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 553
:SBUS<n>:UART:TRIGger:IDLE
(see page 860)
Command Syntax
:SBUS<n>:UART:TRIGger:IDLE <time_value>
<time_value> ::= time from 1 us to 10 s in NR3 format
The :SBUS<n>:UART:TRIGger:IDLE command selects the value of the idle
period for burst trigger in the range from 1 us to 10 s when in UART
mode.
Query Syntax
:SBUS<n>:UART:TRIGger:IDLE?
The :SBUS<n>:UART:TRIGger:IDLE? query returns the current UART trigger
idle period time.
Return Format
<time_value><NL>
<time_value> ::= time from 1 us to 10 s in NR3 format
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:UART:TRIGger:BURSt" on page 551
":SBUS<n>:UART:TRIGger:TYPE" on page 555