:DIGital<d> Commands 12
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 253
:DIGital<d>:THReshold
(see page 860)
Command Syntax
:DIGital<d>:THReshold <value>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<value> ::= {CMOS | ECL | TTL | <user defined value>[<suffix>]}
<user defined value> ::= -8.00 to +8.00 in NR3 format
<suffix> ::= {V | mV | uV}
TTL = 1.4V
CMOS = 2.5V
ECL = - 1.3V
The :DIGital<d>:THReshold command sets the logic threshold value for all
channels in the same pod as the specified channel. The threshold is used
for triggering purposes and for displaying the digital data as high (above
the threshold) or low (below the threshold).
Query Syntax
:DIGital<d>:THReshold?
The :DIGital<d>:THReshold? query returns the threshold value for the
specified channel.
Return Format
<value><NL>
<value> ::= threshold value in NR3 format
See Also "Introduction to :DIGital<d> Commands" on page 247
":POD<n>:THReshold" on page 423
":TRIGger[:EDGE]:LEVel" on page 658
NOTE
This command is only valid for the MSO models.