Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manual de Serviço Página 223

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 970
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 222
:CHANnel<n> Commands 10
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 223
The CHANnel<n> subsystem commands control an analog channel (vertical
or Y- axis of the oscilloscope). Channels are independently programmable
for all offset, probe, coupling, bandwidth limit, inversion, vernier, and
range (scale) functions. The channel number (1, 2, 3, or 4) specified in the
command selects the analog channel that is affected by the command.
A label command provides identifying annotations of up to 10 characters.
You can toggle the channel displays on and off with the
:CHANnel<n>:DISPlay command as well as with the root level commands
:VIEW and :BLANk.
Reporting the Setup
Use :CHANnel1?, :CHANnel2?, :CHANnel3? or :CHANnel4? to query setup
information for the CHANnel<n> subsystem.
Return Format
The following are sample responses from the :CHANnel<n>? query. In this
case, the query was issued following a *RST command.
:CHAN1:RANG +40.0E+00;OFFS +0.00000E+00;COUP DC;IMP ONEM;DISP 1;BWL 0;
INV 0;LAB "1";UNIT VOLT;PROB +10E+00;PROB:SKEW +0.00E+00;STYP SING
NOTE
The obsolete CHANnel subsystem is supported.
Vista de página 222
1 2 ... 218 219 220 221 222 223 224 225 226 227 228 ... 969 970

Comentários a estes Manuais

Sem comentários