Table 8-1. Numbered Error Messages (16 of 17)
22 User tone not within +/-5 dB of -60 dBc.
Press re-try if you wish to re-adjust and continue. The level from the tone
generator is not close enough to the required limits.
Action:
Adjust tone generator if possible. If the tone generator cannot meet this
requirement, replace it with one that can to perform this test.
23 Warning. Measurement has not been calibrated.
'New Msrmnt' should be requested. Repeat Measurement or Noise Monitor was
selected.
Action:
Select New Measurement or New Noise Monitor to allow the system to calibrate
the measurement.
24 No printer found in system configuration, Performance Tests cannot be
initiated.
The Performance Tests require a printer to be connected to the system and
entered in the System Configuration Table.
Action:
Connect a printer and enter the appropriate information into the System
Configuration Table.
25 Printer is not responding. Performance test cannot be initiated. Correct
problem and Press 'Retry'. The system cannot find the printer as addressed in the
System Configuration Table.
Action:
Be sure the printer is connected on the system HP-IB bus. Verify that the printer
is turned on and is set to the same address as is listed in the System Configuration
Table.
26
Tne
system attempted to set (instrument) to (-) MHz. The source was not able to
respond to the frequency setting defined (or computed) for the measurement.
Action:
Verify that the Instrument Parameters defined for the measurement are correct
and that the source's frequency range is adequate.
27 The entered carrier frequency cannot be measured. The HP 11729C cannot
generate an intermediate frequency between 4.9 MHz and 1280 MHz for this carrier
frequency.
Action:
Verify the carrier frequency. If the frequency is correct, verify the HP 11729C filter
table in the System Configuration Table.
Numbered Error Messages 8-17