Description
The Default Table Values
The default RF Segment Table provides an optimum set of values for most
noise measurements from 100 kHz to 40 MHz. The default RF Segment
Table values should be used the first time a new measurement is made.
This guarantees that any measurement problems are not due to the RF
Segment Table.
Plotted
Range (Hz)
100+3 - 500E+3
500+3 - 5E+6
5E+6 - 40E+6
0
0
0
Measurement
Range (Hz)
500E+3
5E+6
40E+6
Min.
Av'g
1
1
1
BW (Hz)
3.E+3
10.E+3
30.E+3
Video
BW (Hz)
10
30
100
Sweep
Time
20
20
20
Storing the Table
Segment table data is stored whenever a Test File is created. When a Test
File is loaded, the RF Segment Table data is set to the values stored in the
Test File.
Modifying the Table
All fields in the RF Segment Table can be user denned, however, the values
entered must reflect available settings for the particular RF analyzer you
are using. The accuracy of the measurement results is dependent on the
values defined in the RF Segment Table.
Changing the table can cause gaps in the Results Graph if the segments
do not cover a continuous frequency range. Also, changes can result in a
longer then necessary measurement time.
Any time the segment table is changed a New Measurement is recom-
mended.
5-24 RF Segmnts