Agilent B1500 Users Guide, Edition 7 2- 41
Introduction
Specifications
HPSMU
HPSMU: High Power Source/Monitor Unit
Voltage range, resolution, and accuracy
See Table 2-19 for high resolution ADC and Table 2-21 for high speed ADC.
Current range, resolution, and accuracy
See Table 2-20 for high resolution ADC and Table 2-22 for high speed ADC.
Specified measurement resolution is limited by fundamental noise limits.
Power consumption
Voltage source mode:
Current source mode:
Voltage range (V)
Power consumption
a
(W)
a. Ic is the current compliance setting.
2 Ic × 20
20 Ic × 20
40 Ic × 40
100 Ic × 100
200 Ic × 200
Volta
ge compliance
a
(V)
a. Vc is the voltage compliance setting.
Power consumption
b
(W)
b. Io is the output current.
Vc 20 Io × 20
20 < Vc 40 Io × 40
40 < Vc 100 Io × 100
100 < Vc 200 Io × 20
0