7- 30 Agilent B1500 User’s Guide, Edition 7
Function Details
SPGU Module
Pulse Output Operation Summary
The SPGU can be used for I/V Sweep, Multi Channel I/V Sweep, I/V List Sweep,
and I/V-t Sampling classic tests. The SPGU pulse output operation will differ as
shown in the following cases. Pulse output example of the sampling measurement is
shown in Figure 7-18.
Sampling measurement, simultaneous output mode. See Table 7-3.
Sampling measurement, sequential output mode. See Table 7-4
Sweep measurement mode. See Table 7-5.
Figure 7-18 SPGU Pulse Output Example for Sampling Measur
ement in Free Run Mode
2: SPGU1
3: SPGU2
Start measurement
base
base
0 V
0 V
base hold time
1: SMU1
base
source
4: SMU2
base
source
0 V
0 V
0 V
0 V
0 V
0 V
3
2
4
1
End of measurement
Simultaneous output:
2: SPGU1
3: SPGU2
Start measurement
base
base
1: SMU1
source
4: SMU2
base
source
0 V
0 V
0 V
0 V
End of measurement
Sequencial output:
base hold time
0 V
0 V
0 V
0 V
3
2
4
1