Agilent B1500 User’s Guide, Edition 7 5- 21
Classic Test Definition
I/V-t Sampling
I/V-t Sampling
On the EasyEXPERT main screen, click the Classic Test tab and select the I/V-t
Sampling icon. The main screen displays the I/V-t sampling test setup screen. For
details of the measurement mode, see “I/V-t Sampling Measurement” on page 7-13.
Setup Name Used to set the name of test setup. The name must be the identification for the test
s
etup save
d to a My Favorite group (preset group). Enter a unique name that is
distinct from other definitions.
The name is recorded in the test result data and is used to different
iate the data in the
test record list area.
Channel Setup This tab displays the Channel Setup screen. See “Channel Setup” on page 5-22.
Measurement
Se
tup
This tab d
isplays the Measurement Setup screen. See “Measurement Setup” on page
5-23.
Function Setup This tab displays the Function Setup screen. See “Function Setup” on page 5-42.
Auto Analysis
Se
tup
This tab
displays the Auto Analysis Setup screen. See “Auto Analysis Setup” on
page 5-43.
Display Setup This tab displays the Display Setup screen. See “Display Setup” on page 5-46.
NOTE If you set Interval < 0.002 s
Sampling mode must be linear. This setting is not permitted for the log sampling.
All measurement channels must use the high speed A/D con
verter (ADC). This
setting is not permitted if a measurement channel uses the high resolution ADC.
If the multiple measurement channels are used, all channe
ls perform measurement
in parallel.
If the measurement ranging mode is not the fixed mode, the meas
urement channels
automatically select the minimum range that covers compliance value set to the
channel.
If the measurement time is expected to be longer t
han Interval, the measurement
channels automatically adjust the number of averaging samples (ADC settings) to
keep the sampling interval.