Agilent B1500 Users Guide, Edition 7 9-57
If You Have a Problem
Error Codes
603 Sweep and pulse channels must be different.
Set the sweep source and the pulse source to differen
t channels for the
staircase sweep with pulsed bias measurement (MM5).
610 Quasi-pulse source channel must be set.
Before triggering the quasi-pulsed spot measurement, send the BDV
command to set the qu
asi-pulse source.
620 TGP specified incorrect I/O port.
Specify trigger input for the Ext Trig In port, or trigger output for the
Ext Trig Ou
t port by the TGP command. See the TGP command of
Programming Guide.
621 Specify trigger input port for PAX/WSX.
No trigger input port was specified for the PAX or WSX command.
Specify the trigge
r input port, or set the port as the trigger input port.
See the TGP command of Programming Guide to set trigger port.
622 Specify trigger output port for OSX.
No trigger output port was specified for the OSX comman
d. Specify the
trigger output port, or set the port as the trigger output port. See the
TGP command of Programming Guide to set trigger port.
630 Incorrect polarity of search step value.
For the linear search measurement. The step value must
be positive if
start<stop, or negative if start>stop.
631 Number of search steps must be 1001 or less.
For the linear search measurement. The number
of search steps
between start and stop must be 1001 or less. This means the |step| value
must be |stop-start|/1001 or more.
632 Search measurement was aborted.
Search measurement was aborted by the automatic ab
ort function.
640 Search limits must be range/20000 or more.
For the binary search measurement. The limit
value for the search
target must be range/20000 or more. where range means the
measurement range actually used for the measurement.
650 Data format must be ASCII to get time data.