2- 16 Agilent B1500 User’s Guide, Edition 7
Introduction
Measurement Units
Measurement Units
Agilent B1500A can be equipped with the following measurement modules.
“GNDU - Ground Unit”
“B1510A High Power SMU”
“B1511A Medium Power SMU”
“B1517A High Resolution SMU”
“B1520A Multi Frequency CMU”
“B1525A High Voltage SPGU”
“B1530A WGFMU”
See “Using EasyEXPERT” on page 4-1 for the functions available for Agilent
EasyEXPERT software. See Programming
Guide for the functions available for the
GPIB remote mode. See Agilent B1530A Users Guide for using the WGFMU.
GNDU - Ground Unit
Agilent B1500A is equipped with the ground unit (GNDU). The GNDU is a 0 V
constant voltage source, and used for the reference of measurement ground. Also the
GNDU can sink up to ± 4.2 A, so it is effective for using the HPSMU (high power
SMU). Figure 2-1 is a simplified circuit diagram of GNDU.
Figure 2-1 Simplified GNDU Circuit Diagram
*1'8
9
)RUFH
&RPPRQ