DisplayPort Test Implementation Group 43 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Scope based ISI measurement for RBR
A horizontal histogram which slices through the center of the crossing point is used to measure jitter
pp
.
The scope is triggered by a signal from J-BERT’s clock or trigger output.
Scope settings:
Set up scope for a measurement on a differential signal either by using a differential SMA probe head
or HW based channel subtraction (channel 1-3 or channel 2-4)
Use 1kpts acquisition depths
Set display mode to color grade
Adjust crossing point to center of the display with two eyes
Activate an horizontal histogram Analyze ! Histogram and select a thin window from the center of
the left eye to the center of the right eye
Activate an histogram peak to peak measurement
J-BERT settings:
Select a clk/2 pattern (1010) on J-BERT for the no ISI measurement and select a PRBS 2
7
-1 pattern
for the pp measurement with ISI
Select 1V
diff
amplitude for the data output
Deactivate SSC and all jitter sources
Set data rate to 1.62Gb/s
For the measurement clear the display and let the scope run for a fixed amount of time, e.g. 2 minutes.
Note the hist
pp
reading and repeat for at least five times. Average the results for the pp no ISI
measurements. Repeat for the hist
pp
with ISI measurements.
Figure 40: RBR ISI measurement step 1 for jitter base line