Agilent TechnologiesAgilent TechnologiesGPIB/USB/LAN InterfacesE2094SIO Libraries Suite 15.5 Connectivity Guidewith Getting Started

Página 2

10 Agilent Connectivity Guide1 Getting StartedUsing This GuideA typical complex test system includes a collection of instruments under the control of

Página 3 - Contents

100 Agilent Connectivity Guide5 Connecting Instruments to LANMultiple NIC Cards If Agilent Connection Expert (see “Step 3: Run Agilent Connection Expe

Página 4 - 4 Agilent Connectivity Guide

Connecting Instruments to LAN 5Agilent Connectivity Guide 1012 Look in the 'Status' column in for 'connected' interfaces.3 Right-c

Página 5 - Agilent Connectivity Guide 5

102 Agilent Connectivity Guide5 Connecting Instruments to LANStep 3: Run Agilent Connection ExpertThis step shows how to • Run Agilent Connection Expe

Página 6 - 6 Agilent Connectivity Guide

Connecting Instruments to LAN 5Agilent Connectivity Guide 103Add and Configure LAN InstrumentsWhen instruments reside on your local area network (LAN)

Página 7 - Agilent Connectivity Guide 7

104 Agilent Connectivity Guide5 Connecting Instruments to LANAgilent Connection Expert automatically locates, identifies, and verifies all LAN instrum

Página 8 - 8 Agilent Connectivity Guide

Connecting Instruments to LAN 5Agilent Connectivity Guide 105Connecting a LAN instrument outside your local subnetWhen an instrument resides outside y

Página 9 - Getting Started

106 Agilent Connectivity Guide5 Connecting Instruments to LANSearching for LAN instrumentsIf you don’t know the IP address or hostname for an instrume

Página 10 - Using This Guide

Connecting Instruments to LAN 5Agilent Connectivity Guide 1071 Enter a range of legitimate IP addresses for Connection Expert to search. In general, k

Página 11 - Agilent Connectivity Guide 11

108 Agilent Connectivity Guide5 Connecting Instruments to LANConnecting a GPIB, USB, etc. instrument on a LAN Gateway1 From the Connection Expert main

Página 12

Connecting Instruments to LAN 5Agilent Connectivity Guide 1092 This opens the LAN instrument configuration dialog box, as shown.Click Find Instruments

Página 13 - Agilent IO Libraries

Getting Started 1Agilent Connectivity Guide 11GPIB• Configure GPIB Network• Install instrument driversSee Chapter 3, Connecting Instruments to GPIB• S

Página 14 - Agilent 488

110 Agilent Connectivity Guide5 Connecting Instruments to LAN3 When the Search for Remote Instruments dialog box appears (as shown below), select GPIB

Página 15 - Agilent Connection Expert

Connecting Instruments to LAN 5Agilent Connectivity Guide 1115 Click OK when you are satisfied that the correct instrument has been located and select

Página 16 - 1 Getting Started

112 Agilent Connectivity Guide5 Connecting Instruments to LANNewly Added LAN Instrument in Connection ExpertFigure 40 Newly Added LAN Instruments in C

Página 17 - VXI Resource Manager

Connecting Instruments to LAN 5Agilent Connectivity Guide 113Figure 41 Change Instrument Properties Dialog Box

Página 18 - Interactive IO

114 Agilent Connectivity Guide5 Connecting Instruments to LANStep 4: Communicate with InstrumentsYou can use the Interactive IO utility within Connect

Página 19 - Agilent VISA Options

Connecting Instruments to LAN 5Agilent Connectivity Guide 115488.2 and SCPI commands for communicating with instruments, as shown below.You can also t

Página 20 - Interactive LXI

116 Agilent Connectivity Guide5 Connecting Instruments to LAN2 Open the Commands menu and select a command. The example below shows a *IDN? command se

Página 21 - Visa Open Reports

Connecting Instruments to LAN 5Agilent Connectivity Guide 1173 Click Send & Read to send the command and receive a response. The results are displ

Página 22 - Contacting Agilent

118 Agilent Connectivity Guide5 Connecting Instruments to LAN5 Run the Microsoft Telnet Client. (On Windows XP, click Start > Run and type telnet 6

Página 23

Connecting Instruments to LAN 5Agilent Connectivity Guide 119Configure a LAN InterfaceWhen you select a LAN interface in the Explorer Pane, the Proper

Página 24

12 Agilent Connectivity Guide1 Getting StartedWhat is Agilent IO Libraries Suite? Agilent IO Libraries Suite is a collection of libraries that give yo

Página 25

120 Agilent Connectivity Guide5 Connecting Instruments to LAN2 Select a LAN interface in the Explorer Pane. Click the Change Properties... button in t

Página 26 - Connection Expert

Connecting Instruments to LAN 5Agilent Connectivity Guide 121Connect timeout. The time, in milliseconds, that the PC will wait when attempting to con

Página 27

122 Agilent Connectivity Guide5 Connecting Instruments to LANTroubleshooting LAN InterfacesThis section shows suggested troubleshooting steps for LAN

Página 28

Connecting Instruments to LAN 5Agilent Connectivity Guide 123plugs into the computer; it should be glowing steadily. If it is not lit or flashes inter

Página 29

124 Agilent Connectivity Guide5 Connecting Instruments to LANAn example to verify communication using the ping command follows.Example: Verifying Comm

Página 30

Connecting Instruments to LAN 5Agilent Connectivity Guide 125.Using the route add Command If you are having difficulty with a private LAN (using a hub

Página 31

126 Agilent Connectivity Guide5 Connecting Instruments to LANWeb Browser ChecksTo see if your PC has network access to an instrument, open the Web bro

Página 32

Connecting Instruments to LAN 5Agilent Connectivity Guide 127Communication Settings ChecksIf all hardware connections are OK and you have been able to

Página 33 - Installation Troubleshooting

128 Agilent Connectivity Guide5 Connecting Instruments to LAN5 Under Proxy server, if “Use a proxy server” is checked, click the Advanced… button6 Und

Página 34 - 34 Agilent Connectivity Guide

Connecting Instruments to LAN 5Agilent Connectivity Guide 1295 Under Check for newer versions of stored pages:, click the “Every visit to the page” ra

Página 35

Getting Started 1Agilent Connectivity Guide 13Agilent IO LibrariesThere are four IO Libraries included in Agilent IO Libraries Suite; each of them all

Página 36

130 Agilent Connectivity Guide5 Connecting Instruments to LANNetwork DiagnosticsMicrosoft provides valuable network diagnostic tools on their TechNet

Página 37

Connecting Instruments to LAN 5Agilent Connectivity Guide 1311 Check Agilent IO Libraries Version. If a version of the Agilent IO Libraries or IO Libr

Página 38

132 Agilent Connectivity Guide5 Connecting Instruments to LANBy default, the Agilent IO Control is always active after the Agilent IO Libraries Suite

Página 39

133Agilent Connectivity GuideAgilent Technologies6Programming Your InstrumentsThis chapter includes:• Programming Overview• Addressing Instruments• Sa

Página 40

134 Agilent Connectivity Guide6 Programming Your InstrumentsProgramming OverviewThis section provides an overview of programming instruments via LAN,

Página 41

Programming Your Instruments 6Agilent Connectivity Guide 135Obtaining Instrument DriversSelected combinations of program environments and I/O are incl

Página 42

136 Agilent Connectivity Guide6 Programming Your InstrumentsUsing Agilent IO Libraries Suite Sample CodeThe Automation-Ready CD with Agilent IO Librar

Página 43 - VISA and VISA COM

Programming Your Instruments 6Agilent Connectivity Guide 137• “VISA COM 488.2 Formatted I/O 1.0" corresponds to BasicFormattedIO.dll2Create Forma

Página 44 - For Executable files:

138 Agilent Connectivity Guide6 Programming Your InstrumentsAddressing InstrumentsThis section gives guidelines to address instruments, including:• Ad

Página 45

Programming Your Instruments 6Agilent Connectivity Guide 139• Readability. Your programs will be much easier to read and understand if, for example, y

Página 46 - 46 Agilent Connectivity Guide

14 Agilent Connectivity Guide1 Getting StartedAgilent SICLAgilent SICL (Standard Instrument Control Library) is an I/O library developed by Agilent th

Página 47

140 Agilent Connectivity Guide6 Programming Your InstrumentsConfigure GPIB Card Parameters. With the GPIB interface selected in the instrument Explore

Página 48 - GPIB Cable

Programming Your Instruments 6Agilent Connectivity Guide 141Example: Addressing GPIB Instruments Using VISA/SICLThis figure shows example VISA and SIC

Página 49 - Install GPIB Cards in Your PC

142 Agilent Connectivity Guide6 Programming Your InstrumentsSince unique names have been assigned by Connection Expert, you can use the VISA viOpen co

Página 50 - Connect GPIB Instruments

Programming Your Instruments 6Agilent Connectivity Guide 1431 Connection Expert automatically configures a USB interface, discovers any USB devices, a

Página 51

144 Agilent Connectivity Guide6 Programming Your InstrumentsLAN Networking ProtocolsThere are two LAN networking protocols provided with the Agilent I

Página 52 - 52 Agilent Connectivity Guide

Programming Your Instruments 6Agilent Connectivity Guide 145• iopen("lan[machineName]:gpib0,1") will use the protocol type configured in Con

Página 53

146 Agilent Connectivity Guide6 Programming Your Instrumentsthe configuration in the LAN Interface Properties Pane on the right. To alter any default

Página 54

Programming Your Instruments 6Agilent Connectivity Guide 147Example: Addressing LAN Instruments Using VISA/SICLThe LAN interface system in this figure

Página 55

148 Agilent Connectivity Guide6 Programming Your InstrumentsSince unique names have been assigned by Connection Expert, you can now use the VISA viOpe

Página 56 - System Controller

Programming Your Instruments 6Agilent Connectivity Guide 149Sample ProgramsSample programs follow to demonstrate sine waveform generation for an Agile

Página 57

Getting Started 1Agilent Connectivity Guide 15Agilent Connection ExpertConnection Expert is a software utility that helps you quickly get your instrum

Página 58

150 Agilent Connectivity Guide6 Programming Your Instruments’ and adjusting the frequency, amplitude, and offset.With Fgen.WriteString "*RST"

Página 59

Programming Your Instruments 6Agilent Connectivity Guide 151Example: Simple Sine Waveform (C# and VISA COM)This program selects the function as “sine,

Página 60 - Instrument

152 Agilent Connectivity Guide6 Programming Your Instruments//// TODO: Add any constructor code // after InitializeComponent call//}/// <summary>

Página 61 - Agilent Connectivity Guide 61

Programming Your Instruments 6Agilent Connectivity Guide 153this.button1.Text = "button1";this.button1.Click += new System.EventHandler(this

Página 62 - GPIB Hardware Checks

154 Agilent Connectivity Guide6 Programming Your Instrumentsprivate void USBInstrument_TextChanged(object sender, System.EventArgs e){}private void bu

Página 63

Programming Your Instruments 6Agilent Connectivity Guide 155Example: Simple Sine Waveform (Visual C++ and VISA)This example program is intended for us

Página 64 - GPIB Software Checks

156 Agilent Connectivity Guide6 Programming Your InstrumentsCHECK(viWrite(Instrument, SCPIcmd, (ViUInt32)strlen(SCPIcmd), &actual));strcpy(SCPIcmd

Página 65 - Check BIOS Setting

Programming Your Instruments 6Agilent Connectivity Guide 157strcpy(SCPIcmd, "FUNCtion:USER PULSE\n"); /* Select the active arb waveform */C

Página 66

158 Agilent Connectivity Guide6 Programming Your Instruments

Página 67 - Interrupt mode for larger

159Agilent Connectivity GuideAgilent Technologies7TCP/IP Network BasicsThis chapter includes:• LAN Interface Overview• TCP/IP Protocols• IP Addressing

Página 68

16 Agilent Connectivity Guide1 Getting StartedFigure 2 Agilent Connection Expert Screen

Página 69

160 Agilent Connectivity Guide7TCP/IP Network BasicsLAN Interface OverviewThis section provides an overview of Local Area Networks (LANs) that use Tra

Página 70

TCP/IP Network Basics 7Agilent Connectivity Guide 161Within the site LAN, every device (PC, router, server, etc.) is called a host. A host is any devi

Página 71 - Connecting Instruments to USB

162 Agilent Connectivity Guide7TCP/IP Network BasicsLAN Hardware ArchitectureA LAN is a way to extend the control of instrumentation beyond the limits

Página 72 - USB Quick Start

TCP/IP Network Basics 7Agilent Connectivity Guide 163or device, the LAN server sends a reply to the LAN interface. This reply contains requested data

Página 73 - USB Interface Overview

164 Agilent Connectivity Guide7TCP/IP Network BasicsFigure 56 Example of Packet-Switched NetworkRouterDataDataDataData DataDataSource HostDestination

Página 74 - Connect USB Instruments

TCP/IP Network Basics 7Agilent Connectivity Guide 165TCP/IP ProtocolsThis section summarizes protocols for LAN networks that use TCP/IP for communicat

Página 75

166 Agilent Connectivity Guide7TCP/IP Network BasicsTCP/IP Layers The TCP/IP network model is based on the Department of Defense (DoD) model. TCP/IP P

Página 76

TCP/IP Network Basics 7Agilent Connectivity Guide 167The Network Interface LayerThe lowest layer in the TCP/IP stack is the Network Interface Layer. T

Página 77

168 Agilent Connectivity Guide7TCP/IP Network BasicsAs packets are sent through the network, each host on the network looks at the packet to see if th

Página 78

TCP/IP Network Basics 7Agilent Connectivity Guide 169Internet Control Message Protocol (ICMP) Internet Control Message Protocol (ICMP) is primarily us

Página 79

Getting Started 1Agilent Connectivity Guide 17Agilent 488Sets Agilent Connection Expert Options. Specifically Agilent VISA options, Configuration Sett

Página 80

170 Agilent Connectivity Guide7TCP/IP Network Basicsdata from the Transport Layer protocols of other hosts. Hosts can be on the same local network or

Página 81 - USB Troubleshooting Overview

TCP/IP Network Basics 7Agilent Connectivity Guide 171sender re-sends the packet. This figure shows a network in which the source and destination are u

Página 82 - 82 Agilent Connectivity Guide

172 Agilent Connectivity Guide7TCP/IP Network Basics• Hypertext Transfer Protocol (HTTP)Ports This figure shows the TCP/IP protocol stack. In the stac

Página 83 - USB Hardware Checks

TCP/IP Network Basics 7Agilent Connectivity Guide 173This figure shows typical socket architecture. When a socket is first created, it has an associat

Página 84 - USB Software Checks

174 Agilent Connectivity Guide7TCP/IP Network BasicsHypertext Transfer Protocol (HTTP) Hypertext Transfer Protocol (HTTP) is an application used to tr

Página 85

TCP/IP Network Basics 7Agilent Connectivity Guide 175IP AddressingThis section describes IP addressing for TCP/IP networks, including:• IP Address Cla

Página 86

176 Agilent Connectivity Guide7TCP/IP Network BasicsExample: Class A, B, and C IP AddressesThis table shows three example IP addresses, where the bold

Página 87

TCP/IP Network Basics 7Agilent Connectivity Guide 177Subnets and Subnet MasksAs noted, an IP address consists of a network portion and a host portion.

Página 88 - 88 Agilent Connectivity Guide

178 Agilent Connectivity Guide7TCP/IP Network BasicsA custom subnet mask adds more bits for the network portion of the IP address and uses fewer bits

Página 89 - Connecting Instruments to LAN

TCP/IP Network Basics 7Agilent Connectivity Guide 179Figure 62 Two Networks (Subnets) Using a RouterLocal Network #2 (4 Hosts)Router(DefaultGateway)Lo

Página 90 - LAN Quick Start

18 Agilent Connectivity Guide1 Getting StartedInteractive IOAgilent Interactive IO is a software utility that allows you to interact with an instrumen

Página 91 - Network Topologies

180 Agilent Connectivity Guide7TCP/IP Network BasicsIP Address Configuration MethodsThis section introduces some IP address configuration methods, inc

Página 92

TCP/IP Network Basics 7Agilent Connectivity Guide 181Dynamic Host Configuration Protocol (DHCP)Dynamic Host Configuration Protocol (DHCP) allows each

Página 93

182 Agilent Connectivity Guide7TCP/IP Network BasicsDuplicate IP Address detection is described by several Internet standards (RFC 2642/RFC 2131/ZEROC

Página 94 - Private Instrument Subnet

TCP/IP Network Basics 7Agilent Connectivity Guide 183To recover from a duplicate IP address condition, eliminate the duplicate IP address condition an

Página 95

184 Agilent Connectivity Guide7TCP/IP Network BasicsDevice Hostname ServicesThis section introduces some device hostname services, including:• Device

Página 96

TCP/IP Network Basics 7Agilent Connectivity Guide 185Dynamic DNS NamingDynamic Domain Name System (Dynamic DNS) is a distributed database of hostnames

Página 97 - Corporate Intranet

186 Agilent Connectivity Guide7TCP/IP Network BasicsRFC NetBIOS NamingRFC NetBIOS Naming is a peer-to-peer naming protocol used by Microsoft File/Prin

Página 98

TCP/IP Network Basics 7Agilent Connectivity Guide 187completes each step in the sequence shown before returning an error message. RFC NetBIOS naming d

Página 99

188 Agilent Connectivity Guide7TCP/IP Network BasicsStatic DNS NamingIn contrast to automatic DNS name resolution, Static DNS naming does not require

Página 100

TCP/IP Network Basics 7Agilent Connectivity Guide 189Configuring Your PC for LAN OperationThis section shows steps to configure your PC, as required,

Página 101

Getting Started 1Agilent Connectivity Guide 19 Table 1 Interactive IO Common CommandsEvent Viewer Invokes the Microsoft Event Viewer, which can be use

Página 102

190 Agilent Connectivity Guide7TCP/IP Network BasicsInstalling Network Interface CardsNetwork Interface Cards (NIC) provide the hardware interface bet

Página 103

TCP/IP Network Basics 7Agilent Connectivity Guide 191Example: Installing TCP/IP (Windows XP)1 Click Start > Network > Internet Connnections. Fro

Página 104

192 Agilent Connectivity Guide7TCP/IP Network BasicsExample: Installing TCP/IP (Windows Vista)1 Click Start > Control Panel > Network Connection

Página 105

TCP/IP Network Basics 7Agilent Connectivity Guide 1934 In the frame, This Connection uses the following items:, select either Internet Protocol Versio

Página 106 - Searching for LAN instruments

194 Agilent Connectivity Guide7TCP/IP Network Basics6 Figure 70 shows an Internet Protocol (TCP/IP) Properties dialog box to automatically obtain IP a

Página 107

TCP/IP Network Basics 7Agilent Connectivity Guide 195Example: Setting PC Hostname (Windows XP)1 Right-click the My Computer icon and then click Proper

Página 108

196 Agilent Connectivity Guide7TCP/IP Network BasicsExample: Setting PC Hostname (Windows Vista)1 Right-click the Computer icon, then click Properties

Página 109

TCP/IP Network Basics 7Agilent Connectivity Guide 197Figure 73 System Properties Screen

Página 110

198 Agilent Connectivity Guide7TCP/IP Network Basics

Página 111

199Agilent TechnologiesAgilent Connectivity GuideGlossaryaddressA string (or other language construct) that uniquely locates and identifies a resource

Página 112

Agilent Connectivity GuideNotices© Agilent Technologies, Inc. 2003 - 2009No part of this manual may be reproduced in any form or by any means (includi

Página 113

20 Agilent Connectivity Guide1 Getting StartedViFind32 (Debug Utility)ViFind32 is a console application that uses the viFindRsrc and viFindNext VISA f

Página 114

200 Agilent Connectivity Guide8 Glossaryacross the street. A bridge examines each message on a LAN, “passing” those known to be within the same LAN, a

Página 115

Glossary 8Agilent Connectivity Guide 201DHCPShort for Dynamic Host Configuration Protocol, a protocol for assigning dynamic IP addresses to devices on

Página 116

202 Agilent Connectivity Guide8 Glossaryhexadecimal characters, sometimes with colon or dash separators between every two characters, such as “00:03:d

Página 117

Glossary 8Agilent Connectivity Guide 203instrumentA device that accepts commands and performs a test or measurement function.instrument driverSoftware

Página 118 - Add a LAN interface

204 Agilent Connectivity Guide8 GlossaryIO LibrariesApplication programming interfaces (APIs) for direct I/O communication between applications and de

Página 119 - Configure a LAN Interface

Glossary 8Agilent Connectivity Guide 205There are many different types of LANs, Ethernets being the most common for PCs. Most Apple Macintosh networks

Página 120

206 Agilent Connectivity Guide8 Glossarylocal-area networks (LANs): The computers are geographically close together (that is, in the same building). w

Página 121

Glossary 8Agilent Connectivity Guide 207FTP: Abbreviation of File Transfer Protocol, the protocol used on the Internet for sending files. HTTP: Short

Página 122 - LAN Troubleshooting

208 Agilent Connectivity Guide8 Glossaryproxy serverA server that sits between a client application, such as a web browser, and a real server. It inte

Página 123 - Using the ping Command

Glossary 8Agilent Connectivity Guide 209each other and configure the best route between any two hosts. Very little filtering of data is done through r

Página 124

Getting Started 1Agilent Connectivity Guide 21Visa Open ReportsVISA Open Reports provide a quick “snapshot” of information about all of the hardware i

Página 125 - from the Command

210 Agilent Connectivity Guide8 Glossaryside-by-sideA side-by-side installation allows two vendors' implementations of VISA to be used on the sam

Página 126 - Web Browser Checks

Glossary 8Agilent Connectivity Guide 211For example, the full address for 150.215.017.009 is 10010110.11010111.00010001.00001001. The Class B network

Página 127 - Communication Settings Checks

212 Agilent Connectivity Guide8 GlossarySystem ControllerOne Controller on a GPIB is the System Controller. This is a master Controller; it has the ab

Página 128

Glossary 8Agilent Connectivity Guide 213to announce its presence on the network to other devices. UPnP devices use Extensible Markup Language (XML) to

Página 129

214 Agilent Connectivity Guide8 GlossaryWindows notification areaSee notification area.

Página 130 - Network Diagnostics

Agilent IO Libraries Suite Getting Started 215IndexNumerics488.2 library, 14Aaddress, 199Address Resolution Protocol (ARP), 168Agilentcontacting Agile

Página 131

216 Agilent IO Libraries Suite Getting Startedgateway IP address, 202getting started, 9glossary, 199GPIBaddressing instruments, 139Agilent IO Librarie

Página 132

Agilent IO Libraries Suite Getting Started 217MMATLAB, 13Media Access Control (MAC) address, 167Microsoft .NET Framework, 31NNational Instruments VISA

Página 133 - Programming Your Instruments

218 Agilent IO Libraries Suite Getting StartedVISA, 12, 13, 213National Instruments, 27Open Report, 15primary, 27, 28secondary, 27, 28side-by-side, 27

Página 134 - Programming Overview

22 Agilent Connectivity Guide1 Getting StartedAgilent Web ResourcesSystem Developer CenterFocus on what matters most: the performance, reliability and

Página 135 - Obtaining Instrument Drivers

23Agilent Connectivity GuideAgilent Technologies2Installing Agilent I/O Libraries SuiteThis chapter includes:• System Requirements for Agilent IO Libr

Página 136

24 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteSystem Requirements for Agilent IO Libraries Suite 15.5The following table lists

Página 137 - Create the New Object. Use

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 25If possible, you should always use the current version of the Agilent IO Librarie

Página 138 - Addressing Instruments

26 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteInstalling the IO Libraries Suite Software This section describes how to install

Página 139

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 274 Insert the Automation-Ready CD with Agilent IO Libraries Suite into the CD-ROM

Página 140

28 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteAgilent VISA is installed as primary. (See “Using Agilent VISA in Side-by-Side M

Página 141 - GPIB (82350) Addressing

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 29If you do have another vendor’s VISA installed, you will see a similar dialog box

Página 142

Agilent Connectivity Guide 3Contents1 Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9Using Th

Página 143

30 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteOn 64-bit Windows systems, you cannot change the destination folders. The Choose

Página 144 - LAN Networking Protocols

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 31• If the Microsoft .NET Framework has not previously been installed on your PC, o

Página 145 - Configuring LAN Interfaces

32 Agilent Connectivity Guide2 Installing Agilent I/O Libraries Suitethere may be another CD with additional software (drivers, sample programs, etc.)

Página 146 - Change Properties

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 33Installation Troubleshooting If you encounter problems while installing the IO Li

Página 147

34 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteSelecting Application SoftwareThis table gives guidelines for using additional I/

Página 148 - TCPIP LAN Client Addressing

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 35Modifying, Repairing, or Removing IO Libraries SuiteOn Windows XP, click Start &g

Página 149 - Sample Programs

36 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteTo Uninstall (remove) or Modify Agilent IO Libraries Suite, you must select Agile

Página 150 - With Fgen

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 37If you see this dialog box, insert the Automation-Ready CD with Agilent IO Librar

Página 151

38 Agilent Connectivity Guide2 Installing Agilent I/O Libraries Suite3 Use the Browse button in the Windows Installer dialog box to browse to the loca

Página 152

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 39This Web site includes many types of instrument drivers, including IVI-COM and VX

Página 153

4 Agilent Connectivity Guide3 Connecting Instruments to GPIB . . . . . . . . . . . . . . . . . . . . . . . . . . .47GPIB Quick Start . . . . . . . . .

Página 154

40 Agilent Connectivity Guide2 Installing Agilent I/O Libraries Suitemay see the message Missing or Damaged GPIB-32.DLL when you attempt to use Nation

Página 155

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 41When you have both Agilent VISA and another vendor's VISA installed on the s

Página 156

42 Agilent Connectivity Guide2 Installing Agilent I/O Libraries Suite• agvisa32.dll is installed, but Agilent's visa32.dll is not. Thus, it will

Página 157

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 43Specifying IVI and VISA Paths in Microsoft Visual StudioIn Visual Studio (VS), yo

Página 158

44 Agilent Connectivity Guide2 Installing Agilent I/O Libraries SuiteFor Include files:<visa_base_path>\WinNT\include<visa_base_path>\WinN

Página 159 - TCP/IP Network Basics

Installing Agilent I/O Libraries Suite 2Agilent Connectivity Guide 45NOTEAgilent IO Libraries version 15.5 installs version 2.0.0 of the IVI Shared Co

Página 160 - LAN Interface Overview

46 Agilent Connectivity Guide2 Installing Agilent I/O Libraries Suite

Página 161 - Private LAN Topology

47Agilent Connectivity GuideAgilent Technologies3Connecting Instruments to GPIBThis chapter includes:• GPIB Quick Start• Step 1: Install Agilent IO Li

Página 162 - LAN Hardware Architecture

48 Agilent Connectivity Guide3 Connecting Instruments to GPIBGPIB Quick StartThis section shows suggested steps to help you quickly get started connec

Página 163 - Packet Switching

Connecting Instruments to GPIB 3Agilent Connectivity Guide 49Step 1: Install Agilent IO Libraries Suite on Your PCBefore you connect your instruments

Página 164 - 7TCP/IP Network Basics

Agilent Connectivity Guide 5USB Troubleshooting Overview . . . . . . . . . . . . . . 81USB Hardware Checks . . . . . . . . . . . . . . . . . . . .

Página 165 - TCP/IP Protocols

50 Agilent Connectivity Guide3 Connecting Instruments to GPIBConnect GPIB Instruments1 Review Connection Guidelines. The recommended method for connec

Página 166

Connecting Instruments to GPIB 3Agilent Connectivity Guide 51Example: Connecting a Single GPIB Instrument The following figure shows connections from

Página 167 - The Network Interface Layer

52 Agilent Connectivity Guide3 Connecting Instruments to GPIBFigure 12 Connecting Three GPIB InstrumentsGPIB CableGPIB Connector.Connect to GPIB porto

Página 168 - The Internet Layer

Connecting Instruments to GPIB 3Agilent Connectivity Guide 53Step 3: Run Agilent Connection ExpertThis step shows how to use Agilent Connection Expert

Página 169 - The Transport Layer

54 Agilent Connectivity Guide3 Connecting Instruments to GPIBWhen you select a particular instrument in the Explorer Pane, you see the parameters of t

Página 170

Connecting Instruments to GPIB 3Agilent Connectivity Guide 554 Configure GPIB Card Parameters. With the GPIB interface selected in the Explorer Pane,

Página 171 - The Application Layer

56 Agilent Connectivity Guide3 Connecting Instruments to GPIB5 Change/Accept the Configuration Values. When the configuration values displayed are acc

Página 172 - 1 2 . . . 65,536

Connecting Instruments to GPIB 3Agilent Connectivity Guide 57Step 4: Communicate with InstrumentsIf your instruments show up in the Explorer Pane as v

Página 173 - 2 65535. . . 1 2 65535. .

58 Agilent Connectivity Guide3 Connecting Instruments to GPIBExample using Interactive IO to control an Agilent 34401A1 Select an instrument in the Co

Página 174 - Locator (URL)

Connecting Instruments to GPIB 3Agilent Connectivity Guide 59488.2 CommandsMeaning*IDN?The Identification query returns manufacturer, model, serial nu

Página 175 - IP Addressing

6 Agilent Connectivity Guide6 Programming Your Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . .133Programming Overview . . . . . . .

Página 176

60 Agilent Connectivity Guide3 Connecting Instruments to GPIBTroubleshooting GPIB InterfacesThis section shows suggested troubleshooting steps for an

Página 177 - Subnets and Subnet Masks

Connecting Instruments to GPIB 3Agilent Connectivity Guide 61Check IO Libraries SuiteInstallation1 GPIB Hardware ChecksCheck cables/connections/pow

Página 178 - Local and Remote Networks

62 Agilent Connectivity Guide3 Connecting Instruments to GPIBGPIB Hardware ChecksThis section gives guidelines to make hardware troubleshooting checks

Página 179

Connecting Instruments to GPIB 3Agilent Connectivity Guide 63Check Device Manager You can use the Windows Device Manager to reinstall the interface ca

Página 180

64 Agilent Connectivity Guide3 Connecting Instruments to GPIBGPIB Software ChecksThis section provides guidelines for GPIB Interface Card software che

Página 181 - AutoIP/ZEROCONF

Connecting Instruments to GPIB 3Agilent Connectivity Guide 65Agilent 488..., then select the check box labeled Enable Agilent GPIB cards for 488 progr

Página 182

66 Agilent Connectivity Guide3 Connecting Instruments to GPIBWhen the IRQ is asserted, the OS calls each ISR in turn until one of them returns TRUE (m

Página 183

Connecting Instruments to GPIB 3Agilent Connectivity Guide 67The default behavior of the 82350 driver is to use Polling mode for transfers of 256 byte

Página 184 - Device Hostname Services

68 Agilent Connectivity Guide3 Connecting Instruments to GPIB• The default formatted IO write buffer size is 128 so when using this default size, form

Página 185 - Dynamic DNS Naming

Connecting Instruments to GPIB 3Agilent Connectivity Guide 69• If the IO icon is displayed, click the icon and click About Agilent IO Control to displ

Página 186 - RFC NetBIOS Naming

Agilent Connectivity Guide 7AutoIP/ZEROCONF . . . . . . . . . . . . . . . . . . . . . . 181Duplicate IP Address Detection . . . . . . . . . . . . .

Página 187

70 Agilent Connectivity Guide3 Connecting Instruments to GPIBIf the blue IO icon is not displayed, either the blue IO icon display has been turned off

Página 188 - Static DNS Naming

71Agilent Connectivity GuideAgilent Technologies4Connecting Instruments to USBThis chapter includes:• USB Quick Start• Step 1: Install Agilent IO Libr

Página 189 - Checking PC Settings

72 Agilent Connectivity Guide4 Connecting Instruments to USBUSB Quick StartThis section shows suggested steps to help you quickly get started connecti

Página 190 - Installing TCP/IP on Your PC

Connecting Instruments to USB 4Agilent Connectivity Guide 73Step 1: Install Agilent IO Libraries Suite on Your PCInstall Agilent IO Libraries Suite an

Página 191 - Install

74 Agilent Connectivity Guide4 Connecting Instruments to USBConnect USB InstrumentsUse the steps in this figure to directly connect USB instruments to

Página 192

Connecting Instruments to USB 4Agilent Connectivity Guide 75Step 3: Run Agilent Connection ExpertThis step provides an overview of how Connection Expe

Página 193 - Setting PC IP Address

76 Agilent Connectivity Guide4 Connecting Instruments to USBAutomatic Identification of USB Devices. When a USB instrument is plugged into the PC, the

Página 194 - Setting PC Hostname

Connecting Instruments to USB 4Agilent Connectivity Guide 77Both a listener check and an identification query are done automatically, as you can see o

Página 195

78 Agilent Connectivity Guide4 Connecting Instruments to USBYou can change this VISA alias name to one of your own choosing before clicking OK. Using

Página 196

Connecting Instruments to USB 4Agilent Connectivity Guide 79Step 4: Communicate with InstrumentsThis step shows how to use Interactive IO to send 488.

Página 197 - TCP/IP Network Basics 7

8 Agilent Connectivity Guide

Página 198

80 Agilent Connectivity Guide4 Connecting Instruments to USB2 Select a command, or enter one in the Command area.3 Click Send & Read to send the c

Página 199 - Glossary

Connecting Instruments to USB 4Agilent Connectivity Guide 81Troubleshooting USB InterfacesThis section shows suggested troubleshooting steps for Unive

Página 200

82 Agilent Connectivity Guide4 Connecting Instruments to USBCheck IO ControloperationUSB Hardware ChecksCheck cableconnections/damageVerify Agilent IO

Página 201 - Ethernet (MAC) address

Connecting Instruments to USB 4Agilent Connectivity Guide 83USB Hardware ChecksThis section gives guidelines to make hardware troubleshooting checks f

Página 202

84 Agilent Connectivity Guide4 Connecting Instruments to USBUSB Software ChecksThis section provides guidelines to make software troubleshooting check

Página 203

Connecting Instruments to USB 4Agilent Connectivity Guide 85Agilent IO Libraries Suite ChecksThis section give troubleshooting guidelines for the Agil

Página 204 - IP address

86 Agilent Connectivity Guide4 Connecting Instruments to USB1Check Agilent IO Libraries Version. If a version of the Agilent IO Libraries or IO Librar

Página 205 - Listener

Connecting Instruments to USB 4Agilent Connectivity Guide 87Check Agilent IO Control Operation When the Agilent IO Libraries Suite was installed, the

Página 206 - protocols follows

88 Agilent Connectivity Guide4 Connecting Instruments to USB

Página 207

89Agilent Connectivity GuideAgilent Technologies5Connecting Instruments to LANThis chapter includes:• LAN Quick Start• Step 1: Install Agilent IO Libr

Página 208

9Agilent Connectivity GuideAgilent Technologies1Getting StartedThis chapter includes:• Using This Guide• What is Agilent IO Libraries Suite?• Agilent

Página 209

90 Agilent Connectivity Guide5 Connecting Instruments to LANLAN Quick StartThis section shows suggested steps to help you quickly get started connecti

Página 210

Connecting Instruments to LAN 5Agilent Connectivity Guide 91Step 1: Install Agilent IO Libraries Suite on Your PCInstall Agilent IO Libraries Suite an

Página 211

92 Agilent Connectivity Guide5 Connecting Instruments to LANIn general:• IP Address: The local router or the corporate DHCP server assigns IP addresse

Página 212

Connecting Instruments to LAN 5Agilent Connectivity Guide 93Directly Connect Instrument to ComputerThis is the simplest possible network topology. One

Página 213

94 Agilent Connectivity Guide5 Connecting Instruments to LANPrivate Instrument SubnetThis configuration creates a private LAN subnet for the instrumen

Página 214 - Windows notification area

Connecting Instruments to LAN 5Agilent Connectivity Guide 95Private Instrument Subnet on Corporate IntranetIn this topology, the computer is on the sa

Página 215

96 Agilent Connectivity Guide5 Connecting Instruments to LANOpen Instrument Subnet of Corporate IntranetThe computer and instruments connect directly

Página 216

Connecting Instruments to LAN 5Agilent Connectivity Guide 97Corporate IntranetFigure 26 Private subnet different from PCIn order for the PC to access

Página 217

98 Agilent Connectivity Guide5 Connecting Instruments to LANAcross the WAN -- Using the InternetFigure 27 Internet access through firewallThis topolog

Página 218

Connecting Instruments to LAN 5Agilent Connectivity Guide 99General Issues with LAN ConfigurationBasic design considerations for a LAN-based test syst

Comentários a estes Manuais