Agilent Technologies, Inc.
DisplayPort Test Implementation Group 25 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Note: Capture a minimum of 1 million unit intervals with the JMD in order to achieve a
repeatable EYE opening measurement. The vertical EYE height shall be measured using
the same methodology as described in B appendix.
Figure 23: J-BERT N4903B HBR2 Eye Height Measurement without Crosstalk
Note: In this example, the measured PJ value is 121mV.
9.4.2 Eye Height Calibration with Crosstalk
1. For HBR2, apply a non-phase aligned crosstalk aggressor signal (D24.3 pattern) with amplitude of
500mVdifferential (1.50V divided by 3) to TP3 as in Figure 4-3. Adjust the amplitude of the crosstalk
aggressor signal until the EYE Height at TP3_EQ is 90mV
diff
.
a. In the J-BERT N4903B go to PG Setup > Clock/Jitter Output menu.
Figure 24: HBR2 crosstalk Amplitude on the J-BERT N4903B