34 Agilent 1100 Series Valves
4 Valve Applications
Alternating Column Regeneration
Alternating Column Regeneration
Alternating column regeneration is a convenient way to increase the sample
throughput. The Agilent 1100 Series 2 position/ 10 port valve (G1157A) can be
used in combination with the Agilent LC and LC/MS modules to increase the
efficiency in laboratories running large amounts of samples.
Gradient elution is a common technique for separation of complex samples in
liquid chromatography, which requires column regeneration before the
subsequent run is started. Using alternating column regeneration valuable
time for the analysis is saved. Core of the alternating column regeneration is
the 1100 Series 2 position / 10 port valve, which allows simultaneous analysis
on one column while a second identical column is flushed and equilibrated.
The schematic setup is displayed in Figure 19.
Figure 19 Alternating Column Regeneration (schematic setup)