24 Agilent 1100 Series Valves
3 Maintenance, Repair, and Troubleshooting
Maintenance and Repair
Replacing the rotor seal for micro valves (G1162A and G1163A)
Disassembling the valve head
1 Use the Hex Key to remove the Stator Screws (1) from the Stator Head (2).
2 Disassemble the Stator Head (2)
3 Remove the three Stator Ring Screws (3) and take off the Stator Ring (4).
4 Remove the Rotor Seal (5) from the Valve Body (6). The Rotor Seal is
mounted on three pins.
Figure 12 micro valve head assembly
1
3
4
5
6
2