Agilent Technologies E6640A EXM Manual de Serviço Página 49

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 340
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 48
Chapter 4 49
Coordinating the Sequencers
Include Source in Sequence is disabled
Coordinating the Sequencers
Once the analyzer list sequencer is initiated, the order of events is as follows:
The controller begins the non-signalling test by instructing the DUT to transmit an RF signal.
The analyzer list sequencer, triggered by the RF Burst received from the DUT, sends a trigger to the
source list sequencer.
The source list sequencer begins executing its sequence and generating an RF output to the DUT.
Vista de página 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 339 340

Comentários a estes Manuais

Sem comentários