Agilent Technologies E6640A EXM Manual de Serviço Página 213

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 340
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 212
Chapter 14 213
W-CDMA Programming Commands
Modulation Accuracy
W-CDMA Programming Commands
17 Total power over a slot (Average) is a floating point number (in dBm) of the
total RF power over a measurement slot.
18 First Slot Number is an integer number of the first slot in Capture Interval. This
is not averaged even if the averaging function is On. It is always the last cycle of
the measurement.
19 DPCCH Slot Format: (floating) If Sync Type is DPCCH, the DPCCH slot
format value used for synchronization is returned.
0.0: Slot Format 0
1.0: Slot Format 1
2.0: Slot Format 2
3.0: Slot Format 3
4.0: Slot Format 4
5.0: Slot Format 5
If Sync Type is PRACH, the value returned is -999.0.
In BTS mode, the value returned is -999.0.
20 Preamble Signature: (floating)
BTS mode:
The returned value is always -999.0.
MS mode:
In Preamble Signature auto-detection mode, the detected signature code number
(from 0.0 to 15.0) is returned when the Sync Type is PRACH Message.
In Preamble Signature manual setting mode, the retuned value is the same as the
parameter setting. When the Sync Type is not PRACH Message, the returned
value is NAN.
21 Average Peak EVM (Average Peak) is a floating point number (in percent) of
the averaging for Peak EVM readings on each slot over the entire measurement
area.
Vista de página 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 339 340

Comentários a estes Manuais

Sem comentários