Agilent Technologies 8590 L-Series Especificações Página 573

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 572
Chapter 13 573
Softkey Descriptions
Softkey Descriptions
CAL TRK GEN Calibrate Tracking Generator
DESCRIPTION Note that this function is available for instruments
equipped with Option 010, 50 Output Tracking
Generator, and Option 011, 75 Output Tracking
Generator, only.
CAL TRK GEN initiates the tracking generator
self-calibration routine.
Be sure to perform the CAL AMPTD and CAL FREQ
self-calibration routines before using CAL TRK GEN.
Connect the RF OUT from the tracking generator to the
RF INPUT using the CAL cable. Refer to Table 1-4 for
the part number of the 50 CAL cable or 75 CAL
cable.
When the
CAL TRK GEN routine is complete, press CAL
STORE tostore the tracking generator correction factors
in nonvolatile memory. The correction factors cannot be
viewed.
CAUTION If the correction factors are not stored, they will be lost when the
analyzer is turned off.
R
ELATED SOFTKEYS CAL AMPTD
CAL FREQ
CAL STORE
Vista de página 572
1 2 ... 568 569 570 571 572 573 574 575 576 577 578 ... 673 674

Comentários a estes Manuais

Sem comentários