192 Chapter 2a
Making Adjustments: If 3335A Source Not Available
15a. Frequency Response of the 8590L, 8591C, and 8591E
Procedure for System Characterization
(75 input)
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz
power sensor as described in the measuring receiver operation
manual.
2. Zero and calibrate the power meter and the 75 power sensor as
described in the power meter operation manual.
3. Press INSTRUMENT PRESET on the synthesized signal generator.
Set the synthesized signal generator controls as follows:
CW .................................................................. 41MHz
FREQ STEP ................................................... 37MHz
POWER LEVEL ............................................... 5dBm
4. Connect the equipment as shown in Figure 2a-4.
Figure 2a-4 System Characterization Test Setup for 75 inputs
CAUTION Use only 75 cables, connectors, or adapters on instruments equipped
with 75 inputs or damage to the input connectors will occur.
5. Adjust the synthesized signal generator POWER LEVEL for a
0 dBm reading on the measuring receiver.