Chapter 13 571
Softkey Descriptions
Softkey Descriptions
Note that the CAL YTF routine must be performed
before running the
CAL MXR routine. New frequency
response correction constants must also be generated
whenever the
CAL MXR routine is used. Refer to the
CAL MXR adjustment routine in Chapter 2 for
instructions on the use of the
CAL YTF and CAL MXR
routines.
Once the
CAL MXR routine is finished, the optimum
bias-current value for each frequency band is displayed.
The displayed values are the mixer bias DAC setting
multiplied by 16. The DAC values are automatically
stored as correction constants in nonvolatile memory on
the A16 processor/video assembly.
R
ELATED SOFTKEYS CAL YTF
EDIT FLATNESS
CAL STORE Calibration Store
DESCRIPTION CAL STORE copies the correction factors from working
RAM to the area of memory on the A16 processor/video
assembly that retains data after the analyzer is turned
off.
CAUTION If the correction factors are not stored, they will be lost when the
analyzer is turned off.
R
ELATED SOFTKEYS CAL AMPTD
CAL FREQ
CAL FREQ & AMPTD
CAL FETCH
CAL TIMEBASE
CRT VERT POSITION
CRT HORZ POSITION
CAL TRK GEN