Chapter 4 241
Troubleshooting the Analyzer
Using the Internal Service-Diagnostic Routines
Verify the ground reference detector
This service-diagnostic routine verifies the ground reference circuitry
on the A16 processor/video board.
Verify the ground reference by pressing
GND REF.
If the signal trace is at the bottom of the display, the ground
reference is functioning properly. See Figure 4-8.
If the signal is not at the bottom of the display, the problem is
isolated to the A16 processor/video board assembly. Refer to
“Troubleshooting the A16 Processor/Video Board Assembly”
located in this chapter for further troubleshooting.
Figure 4-8 Ground Reference Detector