566 Chapter13
Softkey Descriptions
Softkey Descriptions
While CAL: LOG AMP is displayed, the following
calibrations are performed:
The log fidelity is measured in 1 dB steps.
Errors are corrected with digital offsets of the
video signal on the A16 assembly.
The video offsets are stored as correction factors.
If a failure occurs during this calibration routine, an
error message is displayed. Refer to Chapter 14 for the
description of displayed error messages.
Be sure to press
CAL STORE after running the CAL
AMPTD routine.
R
ELATED SOFTKEYS DISPLAY CAL DATA
CAL FREQ
CAL FREQ & AMPTD
CAL FETCH
CAL STORE
CORRECT ON OFF
CAL FETCH Calibration Fetch
DESCRIPTION CAL FETCH retrieves the self-calibration correction
factors from the area of memory that retains data when
the analyzer is turned off and places it in working
(volatile) RAM memory.
CAUTION Pressing CAL FETCH clears the correction-data-error flag and can
permit erroneous data to be stored. Use
CAL FETCH only for
troubleshooting.
R
ELATED SOFTKEYS CAL AMPTD
CAL FREQ
CAL FREQ &AMPTD
CAL STORE