Agilent Technologies 6002A Guia do Utilizador Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 50
The Settings screen 1 Getting Started
E6000C Mini-OTDR Users Guide, E0302
51
NOTE Changes made to the settings screen only affect
subsequent measurement acquisitions. However, you can
apply the changes that you have made to a trace that is
currently running by pressing R
UN/STOP again.
The Measurement Settings screen
The Measurement Settings page contains a list of
parameters that you can set (Figure 12).
Figure 12 The Measurement Parameters Settings page
You can change the following parameters:
Range Range: the start position and the measurement span. If
the Range displays Auto, the OTDR selects a suitable
measurement range for your fiber.
You can choose from one of the predefined ranges, or
select Range Input.. and input a range of your
choice.
Pulse Width PulseWidth: the length of the pulses launched by the
OTDR into the fiber. Short pulses improve resolution,
but longer pulses are required for higher dynamic on
long fibers.
Vista de página 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 319 320

Comentários a estes Manuais

Sem comentários