Agilent Technologies 6002A Guia do Utilizador Página 256

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 255
D Single-Mode/Multimode Module Performance Tests Performance Test Form Sheets
256 E6000C Mini-OTDR Users Guide, E0302
Performance Test for the Agilent E6000C with Multimode Modules Page 5 of 6
Model Agt. __________ Module Report No. __________ Date ___________
No. Test Description
I. Dynamic Range 1300 nm Wavelength
Minimum Specification
Pulsewidth Dynamic
Range
98%
Dyn Range =
Dyn Range
98%
+ 1.9dB
E6005A E6009A Meas.
Uncertainty
10
µs
1
µs
100 ns
10 ns
dB
_______
_______
_______
_______
dB
_______
_______
_______
_______
dB
34
28
22
17
dB
n/a
23
18
12
dB
_______
_______
_______
_______
II. Event Deadzone 1300 nm Wavelength
Event Deadzone Max Spec Meas.
Uncertainty
Return Loss
35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
E6005A, E6009A _______ m 3 m _______ m
Vista de página 255
1 2 ... 251 252 253 254 255 256 257 258 259 260 261 ... 319 320

Comentários a estes Manuais

Sem comentários