Agilent Technologies
EPM Series Power Meters
E-Series and 8480 Series Power Sensors
Technical Specifications
Product Specifications
and Characteristics
E4418B EPM SERIES POWER METER
POWER REF
CHANNEL
Vista de página 0
1 2 3 4 5 6 ... 21 22

Resumo do Conteúdo

Página 1 - Agilent Technologies

Agilent TechnologiesEPM Series Power MetersE-Series and 8480 Series Power SensorsTechnical SpecificationsProduct Specifications and CharacteristicsE44

Página 2 - Power Reference

10The chart in Figure 1 shows the typicaluncertainty in making a relative powermeasurement, using the same powermeter channel and the same powersensor

Página 3 - Supplemental Characteristics

E-series E9300 Average Power Sensor SpecificationsThe E-series E9300 wide dynamic range, average power sensors are designed for use with the EPM famil

Página 4 - Settling Time

12Typical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300A and E9301A sensorsTypical SWR, 9kHz to 6 GHz, (25 ˚C ± 10 ˚C) for E9304A sensors1.181.21.1

Página 5 - For E-series E9300A Sensors

13Typical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300B and E9301B sensorsTypical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300H and E9301H sen

Página 6 - General Characteristics

14Power Linearity (after zero and cal at ambient environmental conditions)Table 10Sensor Power Linearity Linearity(25 ˚C ± 10 ˚C) (0 – 55 ˚C)E9300A, E

Página 7

15E-Series E9300 Average Power SensorsTypical E9300B/01B Power Linearity at 25 ˚C, after zero and calibration, with associatedMeasurement UncertaintyP

Página 8 - 11683A Range Calibrator:

16Figure 2 shows the typical uncertainty in making a relative power measurement, using the same power meter channel andsame power sensor to obtain the

Página 9

17E-Series E9300 Average Power Sensors Switch Point Data:The E9300 power sensors have two paths, a low power path covers: -60 to -10 dBm ("A"

Página 10 - Reflection Coefficient (Rho)

18Calibration Factor (CF) and Reflection Coefficient (Rho)Calibration Factor and Reflection Coefficient data are provided at frequency intervals on a

Página 11

19The 8480 series power sensors aredesigned for use with the 435B, 436A,437B, 438A, 70100A, E1416A and nowthe E4418B and E4419B power meters.Thes

Página 12

2Table 1AccuracyInstrumentationAbsolute: ±0.02 dB (Log) or ±0.5%(linear). Please add the correspondingpower sensor linearity percentagefrom Tables 7 a

Página 13 - Frequency GHz

Model Frequency Range Maximum SWR Power Linearity (1)Maximum Power Connector Type Weight8485A 50 MHz to 26.5 GHz 50 MHz to 100 MHz: 1.15 +10 dBm to +

Página 14

21High Sensitivity Sensors, 100 pW to 10 µW (–70 dBm to –20 dBm)1. Negligible deviation except for those power ranges noted. 2. For pulses greater th

Página 15 - Measurement Uncertainty

For more information about Agilent Technologies test and measurement products, applications, services,and for a current sales office listing, visit ou

Página 16

3Supplemental CharacteristicsPower ReferenceFrequency: 50 MHz nominal.SWR: 1.05 maximum.Connector: Type N (f), 50 ohms.Measurement SpeedThree measurem

Página 17

4Auto filter, default resolution, 10 dB decreasing power step (Normal and x2 modes):For E-Series Sensors:In FAST mode (using Free Run trigger), withi

Página 18 - References:

X2 NormalMode ModeMaximum Sensor Power40 ms 70 msTypicalSensorSettlingDynamicTimesRange-40 dBm70 ms 120 ms-50 dBm210 ms 400 ms-60 dBm3.4 s 6.5 sX2 Nor

Página 19

6Power Meter FunctionsAccessed by key entry: either hardkeys, or softkey menu, and programmable.Zero: Zeros the meter. (Power refer-ence calibrator is

Página 20 - Table 15

71. Characteristics describe product performance that is useful in the application of the product, but is not covered by the product warranty.Battery

Página 21

8Options AvailableOption 001: Supplies internal rechargeable battery.Option 002: Supplies parallel rearpanel sensor input(s). (Power reference calibra

Página 22

Power LinearityTable 7Power Temperature (25 ºC ±5 C) Temperature (0 to 55 ºC)100 pW to 10 mW ±3 % ±7 %(-70 dBm to +10 dBm)10 mW to 100 mW ±4.5 % ±10 %