Test IV. Distance Accuracy (Optional) D Single-Mode/Multimode Module Performance Tests
E6000C Mini-OTDR Users Guide, E0302
239
Figure 83 Attenuation Deadzone Test: Marker A at Start of Reflection
Note result in test record 24 Note the distance between the A-B markers as the
attenuation deadzone in the test record.
Test IV. Distance Accuracy
(Optional)
Connect equipment 1 Connect the equipment as shown in Figure 84.
If you are using an Agilent Recirculating Delay Line,
connect port 2 to the OTDR, and leave port 1 open.