E Cleaning Information How to clean instruments with a fixed connector interface
290 E6000C Mini-OTDR Users Guide, E0302
2 Clean the lens by rubbing the cotton-swab over the
surface using a small circular movement.
3 Using a new, dry cotton-swab remove the alcohol, any
dissolved sediment and dust.
4 Blow away any remaining lint with compressed air.
How to clean instruments with a
fixed connector interface
You should only clean instruments with a fixed connector
interface when it is absolutely necessary. This is because
it is difficult to remove any used alcohol or filaments from
the input of the optical block.
Dust caps It is important, therefore, to keep dust caps on the
equipment at all times, except when your optical device is
in use.
Compressed air If you do discover filaments or particles, the only way to
clean a fixed connector interface and the input of the
optical block is to use compressed air.
Fluids and fat If there are fluids or fat in the connector, please refer the
instrument to the skilled personnel of Agilent’s service
team.
CAUTION Only use clean, dry compressed air. Make sure that the air
is free of dust, water, and oil. If the air that you use is not
clean and dry, this can lead to filmy deposits or scratches
on the surface of your connector interface. This will
degrade the performance of your transmission system.