General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 55
Power Sweep Mode
Power sweep is generally used in power level calibration setup where the
frequency is fixed (CW frequency), and the amplitude of the power source
signal is swept. This mode can be used to characterize the flatness,
linearity or gain compression of a device under test.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN.
Figure 2-19 TRIG IN and TRIG OUT connection diagram between power meter and power
source
3 Press . The Channel Setup screen displays as below.
NOTE
This feature is only available when 8480 Series, N8480 Series, E-Series E4410, or E-Series
E9300 sensor is connected.
TRIG
IN
TRIG
IN
TRIG
OUT
TRIG
OUT
Power Meter
Power Source
Average Power Sensor
(optional connection)