Using 8480 Series Power Sensors 5
N1913/1914A EPM Series Power Meters User’s Guide 137
Procedure
First select the table for the sensor you are using as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and
press to highlight .
Figure 5-6 Sensor table selected
3 Press to complete the selection of the calibration factor table.
Figure 5- 7 shows which offset table is selected.
Figure 5-7 Frequency dependent offset indicator
4 To change the frequency, press and use the and keys to
highlight the Frequency field.
Tab les
Sensor Cal Tables
Tab le
On
NOTE
When no data is contained in the highlighted table, the key is disabled (grayed out).
Tab le
Done
Sensor Table 5 Selected