5 Using 8480 Series Power Sensors
134 N1913/1914A EPM Series Power Meters User’s Guide
10 To change the settings, use the and keys to highlight the Cal
Fac value field and press to display the Cal Factor pop- up. Use the
numeric keypad to enter the required value in the Cal Factor pop- up
window.
Figure 5-4 Calibration factor pop-up window
11 Press
to complete the entry.
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
Figure 5-5 Calibration factor display
%
Calibration Factor