CALibration:STRing <quoted string>
This command records calibration information about your meter. For
example, you can store such information as the last calibration date or
the next calibration due date. The calibration message may contain up to
40 characters. [non-volatile memory]
CALibration:STRing?
This command queries the calibration message and return a quoted
string.
CALibration:VALue <value>
This command specifies the value of the known calibration signal used by
the calibration procedure.
CALibration:VALue?
This command queries the present calibration value.
4
Chapter 4 Remote Interface Reference
Calibration Commands
175