Agilent Technologies 34420A Especificações Página 168

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 167
The standard event register is cleared when:
You send a *CLS (clear status) command.
You query the event register using the *ESR? (event status register)
command.
The standard event enable register is cleared when:
You turn on the power and you have previously configured the meter
using the *PSC 1 command.
You execute a *ESE 0 command.
The standard event enable register will not be cleared at power-on if you
have previously configured the meter using *PSC 0.
Chapter 4 Remote Interface Reference
The SCPI Status Model
166
Vista de página 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 293 294

Comentários a estes Manuais

Sem comentários