CONFigure[:VOLTage][:DC]:DIFFerence
[{<
range
> | AUTO | MIN | MAX | DEF}][,{<
resolution
> | MIN | MAX | DEF}]
This command presets and configures the meter for difference
measurements with the specified range and resolution. This command
does not initiate the measurement. For difference measurements, the
specified range applies to the signal connected to the Channel 1
terminals. Autoranging is selected for voltage measurements on the
Channel 2 terminals.
CONFigure:FRESistance
CONFigure:RESistance
[{<
range
> | AUTO | MIN | MAX | DEF}][,{<
resolution
> | MIN | MAX | DEF}]
This command presets and configures the meter for 2-wire or 4-wire
ohms measurements with the specified range and resolution. This
command does not initiate the measurement.
CONFigure:TEMPerature
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{<
resolution
> | MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{<
resolution
> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{<
resolution
> | MIN | MAX | DEF}]
This command presets and configures the meter for temperature
measurements using the specified probe type and resolution. This
command does not initiate the measurement. DEF defaults to the last
probe type set.
CONFigure?
This command queries the meter’s present configuration and return a
quoted string.
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands