Installation Note 08720-90318 9
Add Option 1D5 to the Firmware (Firmware Revision 6.xx)
IMPORTANT This procedure instructs you in adding Option 1D5 to the firmware with
8719/20/22D network analyzers with an older CPU board and firmware
revision 6.xx. If your firmware revision is 7.xx, refer to Add Option 1D5 to
the Firmware (Firmware Revision 7.xx)” on page 10. If you have firmware
revision is 7.xx, you will have the newer CPU board.
Move the A7 Jumper
1. Move the A7 CC jumper to the ALT position. (Refer to Figure 5.)
Figure 5 A7 CC Jumper Location
Make an Addition to the Displayed Options List
2. Press
.
3. Verify that the analyzer displays ‘OPTION 1D5’’, by pressing
.
Return the A7 Jumper to the Original Position
4. Return the A7 CC jumper to the normal operating position (NRM).
5. Replace the analyzer bottom cover, and all the rear panel guard feet.
6. Continue with “Verify the Frequency Reference Operation” on page 11.
System
SERVICE MENU
PEEK/POKE
PEEK/POKE ADDRESS
5243250
x1
POKE
-1 x1
Preset
System
SERVICE MENU