Chapter 4: Testing Performance
Analog Bandwidth - Maximum Frequency Check
69
9
Press the Clear Display key on the scope and record the scope V rms reading in the Analog
Bandwidth - Maximum Frequency Check section of the Performance Test Record (Vout @
50 MHz). For all scope readings in this procedure, use the mean value in the Measurements
display area at the bottom of the screen.
10 Calculate the reference gain as follows:
Record this value in the Calculated Gain @50 MHz column in the Analog Bandwidth - Maximum
Frequency Check section of the Performance Test Record.
Notes
For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
If a question mark is displayed in front of any of the values at the bottom of the screen, press the Clear
Display key on the scope, wait for the #Avgs value to return to 16 and then record the scope reading.
Record the
mean reading
Gain
50 MHz
V
out @50 MHz
V
in @50 MHz
-------------------------------=