Chapter 4: Testing Performance
DC Gain Accuracy Test
62
up as follows and then click OK:
Source = Channel 1
Measurement Area = Entire Display
5
Press the Clear Display key on the scope, wait for the #Avgs value (top left corner of screen)
to return to 256 and then record the scope's mean V avg reading in the Vscope+ column in the
Offset Accuracy Test (50 W) section of the Performance Test Record.
6 Change the DC Supply voltage to -15 mV.
7 Press the Clear Display key on the scope, wait for the #Avgs value to return to 256 and then
Notes
For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
If a question mark is displayed in front of any of the values at the bottom of the screen, press the Clear
Display key on the scope, wait for the #Avgs value to return to 256 and then record the scope reading.
The question mark indicates that the oscilloscope could not make a reliable measurement. Therefore,
any results that contain a question mark should be considered suspect and you should follow the
directions above.
Record the
mean reading