
Chapter 4: Testing Performance
Offset Accuracy Test
55
5
Change the vertical sensitivity of channel 1 to 10 mV/div, press the Clear Display key, wait
for the #Avgs value (top left corner of screen) to return to 256 and then record the scope V avg
reading in the V
zero-error
column (Offset Accuracy Test tables - both the 50 W table and 1 MW
table) in the Performance Test Record section at the end of this chapter.
6 Repeat step 5 for the remaining vertical sensitivities for channel 1 in the Offset Accuracy Test
tables in the Performance Test Record section.
7 Press Default Setup, then turn off channel 1 and turn channel 2 display on.
8 Configure the scope to measure V avg on Channel 2 as follows:
a Pull down the Setup menu and select Acquisition. When the Acquisition Setup window is
displayed, enable averaging and set the # of averages to 256.
b Change the vertical sensitivity of channel 2 to 5 mV/div.
c Click the V avg measurement icon on the left side of the screen.
d When the Enter Measurement Info window is displayed, ensure that the Vavg function is set up
as follows and then click OK:
Source = Channel 2
Measurement area = Entire Display
9
Press the Clear Display key on the scope, wait for the #Avgs value to return to 256 and then
record the scope’s mean V avg reading in the V
zero-error
column (Offset Accuracy Test tables
- both the 50 W table and 1 MW table) in the Performance Test Record section at the end of
this chapter.
10 Repeat step 9 for the remaining vertical sensitivities for channel 2 in the Offset Accuracy Test
table of the Performance Test Record.
11 Repeat steps 7 through 10 for channels 3 and 4.
12 Ensure that each V
zero-error
is less than the corresponding V
zero-error
Limit entry in the Offset
Accuracy Test table.
Comentários a estes Manuais