2- 28 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
Table 2-9 Resistance Measurement Range
1
Range value Resistance measurement value
Display
resolution
Test
current
2 0 < R 2 1  1A
20 2 < R 20 10  100 mA
200 20 < R 200 100  10 mA
2 k 200 < R 2 k 1 m 1 mA
20 k 2 k < R 20 k 10 m 100 A
200 k 20 k < R 200 k 100 m 10 A
2 M 200 k < R 2 M 1 1 A
20 M 2 M < R 20 M 10 100 nA
200 M 20 M < R 200 M 100 10 nA
1. The resistance measurement range is effective for the resistance measurements set to the AUTO
mode which is selected by the
:SENSe:RESistance:MODE command.