Agilent B2900 SCPI Command Reference, Edition 2 4- 11
Subsystem Commands
:CALCulate:LIMit:COMPliance:FAIL
Query response bit_pattern <newline>
bit_pattern returns the fail bit pattern in the format specified by the
:FORMat:DIGital command. Response data type is NR1 or NDN.
Example :CALC:LIM:COMP:DIG:DATA 64
:CALC2:LIM12:COMP:DIG?
:CALCulate:LIMit:COMPliance:FAIL
Sets the judgement criteria for the compliance status check specified by m.
Syntax :CALCulate[c]:LIMit[m]:COMPliance:FAIL criteria
:CALCulate[c]:LIMit[m]:COMPliance:FAIL?
Parameter criteria OUT|IN (default). Parameter data type is CPD.
criteria=IN determines that the limit test has failed if the channel goes into the
compliance state.
criteria=OUT determines that the limit test has failed if the channel comes out of the
compliance state.
Query response criteria <newline>
criteria returns IN or OUT which indicates the present setting of the judgement
criteria. Response data type is CRD.
Example :CALC:LIM:COMP:FAIL OUT
:CALC2:LIM12:COMP:FAIL?
:CALCulate:LIMit:FAIL?
Returns the result of the limit test specified by m.
Syntax :CALCulate[c]:LIMit[m]:FAIL?
Query response result <newline>
result returns 0 or 1. Response data type is NR1.
0: Passed
1: Failed